SOP - OPERATION AND CALIBRATION PROCEDURE FOR INFRA RED SPECTROPHOTOMETER (MAKE: BUCK SCIENTIFIC)
1.0 OBJECTIVE To ensure that the instrument performs satisfactorily and gives accurate and reproducible results. 2.0 SCOPE This SOP covers operation and calibration procedure for Infra red spectrophotometer and this SOP is applicable to Quality control department. 3.0 RESPONSIBILITY Officer / Executive. 4.0 ACCOUNTABILITY Department Head 5.0 PROCEDURE 5.1 PROCEDURE FOR GENERAL CLEANING 5.1.1 Ensure that the power supply to the instrument is switched OFF before cleaning. 5.1.2 De-dust the instrument with a clean dry cloth every day. 5.1.3 Once in a week remove adhered dust by wet mopping using detergent solution. Afterwards wipe the surface with a clean dry cloth to remove traces of detergent. 5.1.4 Caution to be taken to clean the oven immediately with dry cloth to remove the moisture. 5.2 PROCEDURE FOR OPERATION 5.2.1 Switch on the ...