1.0 | OBJECTIVE : | ||||||
To lay down procedure for operation and calibration of Vankel Dissolution | |||||||
tester, Model VK 7000 with automatic sampling station, Model VK 8000. | |||||||
2.0 | RESPONSIBILITY : | ||||||
2.1 | Quality Control / Analytical Development Officer, whoever is operating the | ||||||
instrument. | |||||||
2.2 | Quality Control Manager to ensure compliance. | ||||||
3.0 | PROCEDURE : | ||||||
3.1 | Operating Instructions | ||||||
3.1.1 | Turn ON the main AC power supply to VK 7000 Dissolution testing station | ||||||
3.1.2 | Turn the power switch ON provided on the left hand side of the VK 7000 drive | ||||||
unit. The red LEDs on the system control will light and the system monitor will | |||||||
display start up screen. | |||||||
VANKEL INDUSTRIES INC. | |||||||
VK 7000 DISSOLUTION TESTER | |||||||
PROGRAM REV 1.27 | |||||||
After a few seconds the previously set values will be displayed on the monitor | |||||||
3.1.3 | Remove the lids on the vessels. Press the 'LIFT D' key twice and release | ||||||
immediately. The drive unit will move to the highest position and stops | |||||||
automatically . | |||||||
Note: To stop the drive unit, press "LIFT Ñ or D " at any position . | |||||||
3.1.4 | Remove one of the aligned vessels from the first row and keep it safely. | ||||||
Fill the water bath with purified water till the water level comes just above the | |||||||
outlet valve from heater / circulator on the left side of the bath. Check for any | |||||||
leaks, tighten the screws on the bulkhead fittings on bath /heater inlet or outlet, | |||||||
if any leak is there. | |||||||
3.1.5 | Replace the vessel in original position. | ||||||
3.1.6 | Switch on the AC power supply to the VK750D Heater/Circulator and ON | ||||||
the power switch on the backside of the unit. Ensure that the monitor display | |||||||
---------- and the heater on indicator glows. | |||||||
3.1.7 | Switch on the power supply to VK 8000 sampling station by keeping the ON/ | ||||||
OFF switch provided at the right hand side in ON position. After initializing | |||||||
display, "READY" will be displayed. | |||||||
3.1.8 | If profile is to be performed, follow the following steps. | ||||||
3.1.8.1 | Press 'MENU' key on VK 8000, Main menu shall be displayed. | ||||||
3.1.8.2 | Press '7' key, '4' key and then '9' key, Sampling head shall move to row no. 9. | ||||||
3.1.8.3 | Remove the tube connections to replacement media/rinse reservoir and | ||||||
remove reservoir, wash with purified water and rinse with media. After | |||||||
washing keep it in original position. | |||||||
3.1.8.4 | Connect the appropriate tubes as marked on the tubes to inlet and outlet of | ||||||
replacement media/rinse reservoir. | |||||||
3.1.8.5 | Fill the replacement media/rinse reservoir with medium. Connect the replace- | ||||||
ment media chamber to replacement media pump with the tube marked 'to | |||||||
replacement media/rinse reservoir'. | |||||||
3.1.8.6 | Press 'H' key, sampling head shall move to home position. Press 'ESC' key | ||||||
thrice. 'READY' shall be displayed on the VK 8000. | |||||||
3.1.8.7 | Remove the media tray of media replacement pump and wash it with purified | ||||||
water and then rinse with media. Fix it in pump. Fill it with media and keep the | |||||||
heater tube in media and close with lid. | |||||||
3.1.8.8 | Immerse the tube marked 'into media tray' in media tray and stick it with | ||||||
cellophane tape on lid of media tray. | |||||||
3.1.9 | Switch on the power supply to peristaltic pump. Put the pump in the forward | ||||||
direction by keeping the FWD/REV switch provided at rear side in FWD | |||||||
position. | |||||||
3.1.10 | Fill purified water in cleaning reservoir upto 1 inch level (approx. 2.5 ltrs) and | ||||||
keep the reservoir on the dissolution vessel table of VK 7000. Fix the filters | |||||||
of required porosity to the cannulas of sampling manifold. | |||||||
3.1.11 | Press 'Ñ LIFT' key and slowly down the drive unit such that the sampling manifold | ||||||
is about 6 inches above the level of media in reservoir. | |||||||
3.1.12 | Press ' Ñ Probes' key and adjust the probes by using Ñ lift key such that the | ||||||
filters fixed to the cannulas are completely immersed in the media. Keep the | |||||||
levers on the peristaltic pump towards left side. | |||||||
3.1.13 | Press the 'MENU' key on VK 8000. 'MAIN MENU' will appear. Press '4' key | ||||||
'ENTER PRIME TIME' will appear. ENTER 99 seconds as prime time using | |||||||
numerical keys and press 'ENTER'. | |||||||
3.1.14 | Press '5' key. 'ENTER PURGE TIME' will appear. ENTER 99 seconds as purge | ||||||
time using numerical keys and press 'ENTER'. Press 'ESC' key. | |||||||
3.1.15 | Check all the cannulas on sampling manifold are properly connected with | ||||||
color coded Teflon tubes. | |||||||
3.1.16 | Press 'CLEAN SYSTEM' key on VK 8000. Press '1' to start cleaning. | ||||||
3.1.17 | After completion of cleaning, 'READY' will be displayed on VK 8000 and | ||||||
Peristaltic pump will off. | |||||||
3.1.18 | Perform the Volume calibration either by manual mode or by Automatic mode. | ||||||
3.1.18.1 | Perform the Volume calibration by Automatic mode as follows. | ||||||
Note : Ensure that the autocalibration fixture is perfectly dry. | |||||||
3.1.18.1.1 | Press 'CAL' key when the system monitor on VK 8000 showing ready. Ensure | ||||||
that the auto calibration fixture is in place and top of the calibration fixture is | |||||||
connected to the rear panel connector marked 'calibration'. Also ensure that | |||||||
pump is ON, in forward direction and levers are towards left. | |||||||
3.1.18.1.2 | Press '2' for auto calibration and system starts calibration. |
3.1.18.1.3 | After calibration is over, VK8000 displays with blinking 'Calibration completed' with | ||||||
alarm. | |||||||
3.1.18.1.4 | Press any key to stop the alarm . Press ESC TO EXIT will be displayed. | ||||||
Remove the autocalibration fixture, empty it, allow it to dry and fix it in original | |||||||
position and Press 'ESC' key. | |||||||
3.1.18.1.5 | Verify the calibration by giving a dummy run as follows. | ||||||
3.1.18.1.6 | Press PROG key, and write a programme for ROW '0' with sample time points at | ||||||
02 minutes and R-M Option in DISABLE condition. | |||||||
3.1.18.1.7 | Keep six 10 ml calibrated test tubes in '0' Row in 1 to 6 columns. Press 'START | ||||||
PROG' key and enter the program number. Press '1' key three times. | |||||||
3.1.18.1.8 | System gives calibrated volume in to the test tubes kept in '0' row. Then the system | ||||||
goes for purging. After purging the system comes to READY position. | |||||||
3.1.18.2 | Perform the Volume calibration by manual mode as follows. | ||||||
3.1.18.2.1 | Ensure that the peristaltic pump is ON, in forward direction and levers are | ||||||
towards left. | |||||||
3.1.18.2.2 | Press 'CAL' key when the system monitor on VK 8000 showing ready. | ||||||
Remove the autocalibration fixture, Media/Rinse reservoir and sample tray. | |||||||
3.1.18.2.3 | Press '1' for Manual calibration and system starts calibration. | ||||||
After priming is completed, the sampling valves comes down. | |||||||
3.1.18.2.4 | Place a Calibrated 10 ml test tube or volumetric flask below the valve number 1. | ||||||
Press OPEN VALVES key and the valve start releasing the water. | |||||||
3.1.18.2.5 | Press OPEN VALVES key again when exactly 10 ml of water is collected. | ||||||
Similarly perform the calibration for other valves from 2 to 6. | |||||||
Then valves moves up and the system starts purging. | |||||||
3.1.18.2.6 | After purging, Sampling station moves forward. VK8000 displays with blinking | ||||||
Calibration completed' with alarm. Keep the Auto calibration fixture and | |||||||
media/rinse reservoir in place. |
3.1.18.2.7 | Press any key except ESC key to stop the alarm. 'PRESS ESC TO EXIT' is | ||||||
displayed on the screen. Press 'ESC' key. System goes back to home position. | |||||||
Keep the sample tray in place. | |||||||
3.1.18.2.8 | Repeat operation mentioned in point no.3.1.18.1.5 to 3.1.18.1.8. | ||||||
3.1.19 | Remove the cleaning reservoir from the table carefully by raising up the drive | ||||||
unit,empty it and fill the tray with media upto 1 inch level(approx. .2.5 ltrs.) and | |||||||
follow point nos.3.1.10 to 3.1.15. | |||||||
3.1.20 | After completion of cleaning, 'READY' will be displayed on VK 8000 and | ||||||
Peristaltic pump will off. | |||||||
3.1.21 | Remove the cleaning reservoir from the table carefully by raising up the drive | ||||||
unit and keep it aside. Fill the specified quantity of medium in vessels | |||||||
mentioned in the individual monograph. | |||||||
3.1.22 | Press "MENU" key on VK 7000 to get into the main menu. | ||||||
3.1.23 | Press "2" to set clock. 'ENTER DATE' will be displayed with the current date. | ||||||
Press "ENTER" to accept it, otherwise key in the new date using 0-9 numeric | |||||||
keys in mm/dd/yy format. Press 'ENTER' key. | |||||||
3.1.24 | Now 'ENTER TIME :' with current time of the day will be displayed in 24Hr. format | ||||||
Press "ENTER" to accept the displayed time otherwise key in the new time in | |||||||
24 Hr. format. Press 'ENTER' key. | |||||||
3.1.25 | Press "3" to set temperature '** * SELECT TEMPERATURE SET MODE * * * | ||||||
will be displayed. | |||||||
3.1.26 | Press "1" , the previously set bath temperature will be displayed. Press | ||||||
'ENTER' to accept it, otherwise enter, the exact temperature required (37.5°C) | |||||||
and press 'ENTER'. | |||||||
Note : If a mistake is made, press "CLEAR" the previous entry will be displayed again. | |||||||
3.1.27 | Press '4'. '*** REPORT CENTER CONTROL ***' will be displayed. | ||||||
3.1.28 | Press '1' to keep printer ON. Press '4' to again get into Report center control. | ||||||
Press '4' to ENTER print out frequency. ENTER print out frequency in minutes using | |||||||
0 to 9 keys. Press 'ENTER' key. |
3.1.29 | Press '5' to set print out ID. ENTER the numerical of the instrument ID No. and | ||||||
press 'ENTER' (for example 080 for QAE080). | |||||||
3.1.30 | Press 'MENU' key on VK 7000 to get into Main menu. Press '5' to write the | ||||||
program. '*** SELECT BASKET OR PADDLE' will appear. | |||||||
3.1.31 | Select baskets by pressing '1' or paddles by pressing '2' which ever is | ||||||
mentioned in the monograph and then press MENU key. | |||||||
3.1.32 | Fix the selected ones as described below. | ||||||
3.1.32.1 | Glide each shaft up into the spindle so that all the uncoated part of the paddle | ||||||
shaft goes into the spindle clutch or insert the basket holder shaft with | |||||||
basket into the clutch spindle until the shaft is seen from clutch opening. | |||||||
3.1.32.2 | Tighten the clutch screw on the spindle. | ||||||
3.1.32.3 | Loosen the thumb screw on stop ring and remove the paddle/ basket height | ||||||
gauge from the left second leg and slide the stop ring down completely and | |||||||
tight the thumb screw. | |||||||
3.1.32.4 | Press 'LIFT Ñ' and hold to lower the drive head, hold the key until the drive | ||||||
head stops automatically. | |||||||
3.1.32.5 | Loose the clamp on each spindle and carefully lower each shaft until the | ||||||
paddle blade or basket is resting on the bottom of the vessel. Tight each | |||||||
clamp. | |||||||
3.1.32.6 | Press and hold the 'LIFT D' key and release the key after the drive unit raises 3 | ||||||
or 4 inches up. Loose the thumb screw on stop ring. Slide it up and clip the | |||||||
paddle/basket height gauge on the leg between the stop ring and safety | |||||||
stop. Slide the gauge down so that it rests on the safety stop. Then slide the | |||||||
stop ring down until it rests on the top of the gauge. Tight the stop ring thumb | |||||||
screw. | |||||||
3.1.32.7 | Ensure that appropriate depth spacer (height gauge) is used for basket / | ||||||
paddle whichever is fixed |
3.1.32.8 | Press and hold the 'LIFT Ñ key until the drive unit stops automatically. | ||||||
3.1.33 | Keep the lids on each vessel. | ||||||
3.1.34 | Press 'MENU' key on VK 7000 and then press '5' key and ensure the * mark is | ||||||
on selected option in select basket or paddle menu. Press 'ENTER' key. | |||||||
3.1.35 | 'ENTER RPM' will be displayed. ENTER the RPM value specified in the monograph | ||||||
using 0-9 keys and press 'ENTER' key. | |||||||
3.1.36 | 'ENTER SET BATCH TEMPERATURE' will be displayed. ENTER 37.5°C as bath | ||||||
temperature and press 'ENTER' key. | |||||||
3.1.37 | 'ENTER TEST LENGTH :' shall be displayed. Enter test length 2 minutes more | ||||||
than the mentioned time in individual monograph using 0-9 numerical keys in | |||||||
HH:MM format and press 'ENTER' key. | |||||||
Ex.: Enter 32 minutes test length for 30 minutes. | |||||||
3.1.38 | Final spin at 250 RPM will be displayed. ENTER the final spin time to be at 250 | ||||||
RPM in mm:ss format as per monograph, if required. Press 'ENTER' key. | |||||||
3.1.39 | SAMPLE INITIAL VESSEL TEMPERATURE | ||||||
1. Yes 2. No. Will be displayed. Press 2 key. | |||||||
3.1.40 | SAMPLE FINAL VESSEL TEMPERATURE | ||||||
1. Yes 2. No. Will be displayed. Press 2 key. | |||||||
3.1.41 | 'SAMPLE MANIFOLD AUTO CONTROL' will be displayed. Press '2' to disable. | ||||||
3.1.42 | Ensure that remote interface cable is connected to start input of VK8000 | ||||||
and start output of VK 7000. | |||||||
3.1.44 | Press 'PROG' key on VK 8000. ENTER Prog#(1-30):_____ will be displayed with | ||||||
cursor flashing at entry space. | |||||||
3.1.45 | Enter the program number (from 1-30 only) and press 'ENTER' key. | ||||||
3.1.46 | Program variables menu will be displayed on VK 8000. |
3.1.47 | Press '1' *** Set VK 7000 Parameters *** will be displayed. | ||||||
1. Set COM Port ID. | 4. Final Spin length. | ||||||
2. Set spindle RPM. | 5. Set Printer. | ||||||
3. Set Bath temperature. | 6. Set DDM or Basket. | ||||||
3.1.49.1 | Press 1 to set com port ID No. Keep 01 always. And press ENTER key. | ||||||
3.1.49.2 | Press 2 key two times to set RPM as specified in individual monograph. And press | ||||||
ENTER key. | |||||||
3.1.49.3 | Press 3 key and 2 key to set VK7000 temperature (Bath temperature). Use 0-9 keys | ||||||
to feed the bath temperature and press ENTER key. | |||||||
3.1.49.4 | Press 4 to set final spin length in minutes : Seconds format and press ENTER key. | ||||||
3.1.49.5 | Press 5 to set printer frequency at VK7000 in minutes and press ENTER key. | ||||||
ENABLE OR DISABLE VK7000 PRINTER | |||||||
1. Enable 2. Disable will be displayed. Press '1' key. | |||||||
3.1.49.6 | Press 6 to select Basket or paddle. | ||||||
DOSAGE DELIVERY MODE OF VK7000' | |||||||
1. Sequential 2. Simultaneous will be displayed. Press '1' key. | |||||||
3.1.49.7 | Press 'ESC' key to download set parameters to VK7000? "Y/N" appears, press "Y". | ||||||
***PROGRAMME VARIABLES*** will be displayed. | |||||||
3.1.50 | Press '2' sample time point will be displayed. | ||||||
3.1.51 | Enter the sampling time as per monograph in hh:mm format. E.g.. to take | ||||||
sample after 45 min enter 000:45. Press 'ENTER'. Row number will change to | |||||||
'1'. Enter 000:00. Press 'ENTER' to abort sampling after given time. | |||||||
3.1.52 | To perform dissolution profile with particular time interval gap in sampling | ||||||
enter the next sampling point also. | |||||||
e.g.: To draw samples at an interval of 5 min upto 30 min, Enter sampling time | |||||||
as 000:05 at row '0'. Press Enter and 000:10 at row '1'. Press Enter key and | |||||||
000:15 at row '2'. Press Enter and 000:20 at row '3'. Press Enter key and | |||||||
000:25 at row '4'. Press Enter and 000:30 at row '5'. Press Enter. Row number will | |||||||
change to '6'. Enter 000:00. Press 'ENTER' twice to abort sampling after given time. |
3.1.53 | Again program variables will appear. Press '3' to enter sample volume. Enter | ||||||
sample volume in ml (upto 14 ml only). Press 'ENTER' key. | |||||||
3.1.54 | Press '4' to ENTER batch selection. Press '1' to enter operator name (10 | ||||||
characters only). Press 'ENTER'. Press '2' to enter product name (10 characters | |||||||
only). Press 'ENTER'. Press '3' to enter Batch number (10 characters only). Press | |||||||
'ENTER'. Press '4' to enter lot number (10 characters only). Press 'ENTER'. Press | |||||||
'5' to enter group number (10 characters only). Press 'ENTER'. Press '6' to enter | |||||||
notation (10 characters only). Press 'ENTER' key. | |||||||
3.1.55 | Press 'ESC' to enter into program variables. Press '5' to enter the prime time | ||||||
(enter from 00 to 99 seconds only). Press 'ENTER'. | |||||||
3.1.56 | Press '6' to enter purge time. Enter upto 99 seconds only. Press 'ENTER' key. | ||||||
3.1.57 | Press '7' and press '1' to disable media replacement option or '2' to enable | ||||||
media replacement. It will ask R-M percentage, enter 100 & enter for 100% | |||||||
replacement. | |||||||
3.1.58 | Ensure that the levers on peristaltic pump are towards left side. The test tubes | ||||||
in sample tray are cleaned and dried. | |||||||
3.1.59 | Press 'ESC' key. READY will be displayed. | ||||||
3.1.60 | Ensure that remote interface cable is connected to start input of VK8000 | ||||||
and start output of VK 7000. | |||||||
3.1.61 | Ensure that required number of clean & dry test tubes are placed in sample | ||||||
tray of VK8000. | |||||||
3.1.62 | Ensure that the temperature in all vessels is 37 ± 0.5°C by keeping Thermometer | ||||||
or Sensor of digital thermometer or sensor of QA station in each vessel. Record | |||||||
the initial temperature of each vessel in record of analysis. | |||||||
3.1.63 | Drop tablets/capsules in each vessel and immediately press 'Start Program' on | ||||||
VK8000 for apparatus -2. For apparatus -1, place 1 tablet/capsule in each | |||||||
basket, fix baskets to shafts and down the drive unit until it stops automatically. | |||||||
Press Start program on VK8000, It will ask "Enter Prog (1-30)", Enter the program | |||||||
which you want to run "Press Enter". "Select Start Mode" submenu will be displayed, | |||||||
press 1 to select "Start Now" option. VK7000 options submenu will be displayed, | |||||||
press 1 to select "Load" and "Run" VK7000. Select VK7000 Program start mode, | |||||||
Press 1 to select "Instant", then the run starts. |
3.1.63 | When the particular row sampling is completed the instrument gives a beep | ||||||
sound, and flashes the message 'Current row sampling is completed'. Press | |||||||
any key (except ESC) to proceed with further sampling. | |||||||
Note : If you want to abort the running program press 'ESC' and 'Y'. | |||||||
3.1.64 | After the execution of complete program 'printing....... please wait......' will be | ||||||
displayed with a current row sampling completed alarm. Press any key after | |||||||
printing is over. Ready will be displayed. | |||||||
3.1.65 | Analyse the samples as mentioned in individual monograph. | ||||||
3.1.66 | Remove the lids over vessels, press 'lift D' twice in a second to raise up the | ||||||
drive units. Clean the dissolution apparatus by following SOP "Cleaning of Dissolution | |||||||
Apparatus". | |||||||
3.1.67 | ENTER the usage details in "Instrument/Equipment user log book" (Annexure -4) | ||||||
3.2 | Calibration of RPM, Wobble, Shaft alignment, PT 100 Sensor and | ||||||
Temperature distribution. | |||||||
3.2.1 | Calibrate by following the SOP on "Operation and Calibration of Dissolution | ||||||
QA Station". | |||||||
3.2.2 | Acceptance Criteria : | ||||||
Parameter | Criteria | ||||||
RPM | Should not vary by ±1 of the set value. | ||||||
Wobble | Should not be more than 1 mm | ||||||
Shaft alignment | Deviation should be not more than 1.5° on both | ||||||
x-axis and y-axis. | |||||||
Temperature | The temperature shown by all the eight | ||||||
vessels shall be 37± 0.5°C. | |||||||
3.2.3 | Record the observations in the calibration record (Annexure 1). | ||||||
3.2.4 | Calibration schedule : Once in 3 months or after each major maintenance job. | ||||||
3.3 | Calibration of Timer : | ||||||
3.3.1 | Operate the equipment as mentioned in point No. 3.1.1 to 3.1.4, and 3.1.32 | ||||||
to 3.1.39 by Entering test length as 30 min. | |||||||
3.3.2 | Press 'RUN' key and start the calibrated stopwatch simultaneously. | ||||||
3.3.3 | At the end of 30 minutes, immediately stop the stopwatch. | ||||||
3.3.4 | Record the set time and the time shown by calibrated stopwatch in the | ||||||
"Dissolution apparatus calibration record" (Annexure -2). | |||||||
3.3.5 | Acceptance Criteria : The difference between set time and the time shown | ||||||
by calibrated stopwatch shall not be more than 36 seconds. | |||||||
3.3.6 | Calibration schedule : Once in 3 months or after each major maintenance job. | ||||||
3.4 | Dissolution Apparatus Suitability test by using Prednisone tablets RS | ||||||
and salicylic acid tablets RS : | |||||||
3.4.1 | Proceed as directed in the General Test Procedures "Dissolution Apparatus- | ||||||
suitability test by using Prednisone tablets RS " and "Dissolution Apparatus- | |||||||
suitability test by using Salicylic acid tablets RS ". Record the data in "Worksheet - | |||||||
Calibration Record". For Prednisone tablets in "Suitability test with USP "Prednisone | |||||||
tablets" (Annexure-5). For Salicylic acid tablets in "Suitability test with USP | |||||||
Salicylic acid tablets" (Annexure-6). Get these "Worksheet - Calibration Records" | |||||||
from documentation cell. | |||||||
3.4.2 | Record the results in format "Dissolution Apparatus - Suitability test" | ||||||
separately. (Annexure -3) | |||||||
3.4.3 | Acceptance Criteria : The results shall conform with the specifications | ||||||
provided by USP Convention Inc., for the particular Lot No. of Prednisone | |||||||
tablets RS and Salicylic acid tablets RS. | |||||||
3.4.4 | Calibration Schedule : Once in 6 months or after each major maintenance job. |
3.5 | Calibration of media sampled and replaced : | ||||||
3.5.1 | Perform the volume calibration by following point no. 3.1.18.2 to 3.1.20. | ||||||
3.5.2 | Clean the system for three times by priming and purging for 99 secs with purified | ||||||
water. | |||||||
3.5.3 | Put off the heater and ensure that the instrument is at ambient temperature. | ||||||
3.5.4 | Take about 80 mL of purified water into six LOD bottles and weigh them (W1) | ||||||
Keep them below the probes. | |||||||
3.5.5 | Weigh six test tubes and keep in '0' row of sampling collection tray (W2). | ||||||
3.5.6 | Perform a dummy run with a run time of 2 mins by keeping media replacement | ||||||
option in 'enable' mode and sampling volume 10 mL. | |||||||
3.5.7 | After completion of the run, weigh sample collected in the test tubes (W3) and | ||||||
weigh the LOD bottles with water (W4). | |||||||
Calculate the media replaced by using the following formula : | |||||||
Media replaced (L2) : (W4-W1) + (W3-W2) g | |||||||
Media replaced (L2) : (W4-W1) + (W3-W2) x 1.00396 mL | |||||||
Media sampled (L1) : (W3- W2) g | |||||||
Media sampled (L1) : (W3- W2) x 1.00396 mL | |||||||
Record them in calibration record for media replacement (Annexure-7). | |||||||
3.5.8 | Calculate the difference between media replaced and media sampled for all the six | ||||||
channels. | |||||||
3.5.9 | Acceptance Criteria : The difference between media sampled and replaced should | ||||||
be not more than 0.7 mL and the media sampled should be 10 ± 0.7 mL. | |||||||
3.5.10 | Calibration schedule : Once in 3 months and after any major maintenance job. |
4.0 | DOCUMENTATION : | ||||||
4.1 | Specimen format of "Calibration record" for Calibration of RPM, Wobble, | ||||||
Shaft alignment, temperature Sensors and temperature distribution - Annexure 1. | |||||||
4.2 | Specimen format of "Calibration record" for calibration of Timer - Annexure 2. | ||||||
4.3 | Specimen format of "Dissolution apparatus - suitability test " - Annexure 3. | ||||||
4.4 | File the calibration records in Dissolution calibration record file, kept near the | ||||||
Dissolution apparatus. | |||||||
4.5 | Specimen format of "Instrument/Equipment user log book" - Annexure 4. | ||||||
4.6 | Specimen format of "Dissolution Apparatus - Suitability test with USP Prednisone | ||||||
tablets" worksheet - Annexure - 5. | |||||||
4.7 | Specimen format of "Dissolution Apparatus - Suitability test with USP Salicylic Acid | ||||||
Tablets" worksheet - Annexure - 6. | |||||||
4.8 | Specimen format of "Calibration record for media replacement" - Annexure 7 |